产品列表PRODUCTS LIST

首页>产品中心>日本光明北川式气体检测管>气体快速检测管>141SB二硫化碳CS2气体检知管141SB CARBON DISULPHIDE 0.8-50ppm

二硫化碳CS2气体检知管141SB CARBON DISULPHIDE 0.8-50ppm

简要描述:

二硫化碳CS2气体检测管141SA CARBON DISULPHIDE 0.8-50ppm 30-500ppm,141SB CARBON DISULPHIDE 二硫化碳CS2气体检知管141SB CARBON DISULPHIDE 0.8-50ppm,检测试剂是通过在硅胶或氧化铝等物质表面附着显色试剂的方法制成的。显色试剂遇到被检测气体时会发生特定的反应,而且这种检测试剂具有相当的稳定性。

更新时间:2024-10-06

发邮件给我们:554328895@qq.com

分享到: 1
在线留言
二硫化碳CS2气体检知管141SB CARBON DISULPHIDE 0.8-50ppm

二硫化碳CS2气体检知管141SB CARBON DISULPHIDE 0.8-50ppm

Tube 41SB
1. PERFORMANCE
1) Measuring range :2-50 ppm 0.8-20 ppm
Number of pump strokes 2(200ml) 4(400ml)
2) Sampling time :4 minutes/2 pump strokes
3) Detectable limit :0.3 ppm(400mr)
4) Shelf life :3 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 2 pump strokes.
8) Colour change :Pink→Yellow
2. RELATIVE STANDARD DEVIATION
C.V-low:10% C.V.-mid.:10% C.V.-high:10%
3. CHEMICAL REACTION
Sulphur dioxide is produced by an Oxidizer.
By reacting between this Sulphur dioxide and alkali,
PH indicator is discoloured.
CS2+CrO3+H2SO4→SO2
SO2+2NaOH→Na2SO3+H2O
4. CALIBRATION OF THE TUBE
STANDARD GAS CYLINDER METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
5
2
10
20
30
40
50
ppm
200ml
GAS To pump
(White) (Yellow)
PRETREAT TUBE DETECTOR TUBE
141SB CARBON DISULPHIDE
Substance ppm Interference ppm Coexistence
Sulphur dioxideFIG.1 15 Similar stain is produced. Higher readings are given.
Hydrogen sulphiFdIeG.2 100 〃120 〃
Chlorine Pale pink stain is produced. 〃
(NOTE)
In case of 4 pump strokes, following formula is available for the actual concentration.
Actual concentration = 2/5×Temperature corrected value
0
10
20
30
40
50
5
Sulphur dioxide〔ppm〕�
FIG.1 Influence of Sulphur dioxide
Carbon dioxide
indicating value 〔ppm〕
20 40
Coexistence
Interference
0
10
20
30
40
50
5
Hydrogen sulphide〔ppm〕�
FIG.2 Influence of Hydrogen sulphide
Carbon dioxide
indicating value 〔ppm〕
100 200
Coexistence
Interference
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
50 72 59 50 43 38
45 65 54 45 39 34
40 58 48 40 34 30
35 51 42 35 30 26
30 44 36 30 26 22
25 37 30 25 21 18
20 31 25 20 17 14
15 24 19 15 13 10
10 16 12 10 8 7
5 9 7 5 4 3
2 3 3 2 2 1
www.r-biopharm.com.cn