首页>产品中心>日本光明北川式气体检测管>气体快速检测管>135SA四氯乙烯Cl2C=CCl2气体检测管135SA TETRACHLOROETHYLENE 5-300p

四氯乙烯Cl2C=CCl2气体检测管135SA TETRACHLOROETHYLENE 5-300p

简要描述:

四氯乙烯Cl2C=CCl2气体检测管135SA TETRACHLOROETHYLENE 5-300ppm,中文名称2: 全氯乙烯,英文名称2: perchloroethylene,健康危害: 本品有刺激和麻醉作用。吸入急性中毒者有上呼吸道刺激症状、流泪、流涎。随之出现头晕、头痛、恶心、运动失调及酒醉样症状。口服后出现头晕、头痛、倦睡、恶心、呕吐、腹痛、视力模糊、四肢麻木,甚至出现兴奋不安、抽搐乃至昏迷,可致死。慢性影响:有乏力、眩晕、恶心、酩酊感等。可有肝损害。皮肤反复接触,可致皮炎和湿疹。

更新时间:2024-04-24

发邮件给我们:554328895@qq.com

分享到: 1
在线留言
四氯乙烯Cl2C=CCl2气体检测管135SA TETRACHLOROETHYLENE 5-300p

四氯乙烯Cl2C=CCl2气体检测管135SA TETRACHLOROETHYLENE 5-300ppm

1. PERFORMANCE
1) Measuring range :10-300 ppm 5-150 ppm
Number of pump strokes 1/2(50ml) 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :1 ppm(100ml)
4) Shelf life :2 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Red
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By decomposing with an Oxidizer, Hydrogen chloride is produced and PH indicator is discoloured.
CCI2=CCI2+PbO2+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
150
100
100
30
20
10
5
ppm
GAS To pump
PRETREAT REAGENT
(Grey)
Tube No.
135SA TETRACHLOROETHYLENE
Substance Interference Coexistence
Vinyl chloride Similar stain is produced. Higher readings are given.
Hydrogen chloride FIG.1 〃〃
1,2-Dichloroethylene FIG.2 〃〃
Trichloroethylene 〃〃
Chlorine Pale red stain is produced.
(NOTE)
In case of 1 /2 or 2 pump strokes, following formula is available for the actual concentration.
Actual concentration=2×Temperature corrected value
0
200
100
Hydrogen chloride〔ppm〕�
FIG.1 Influence of Hydrogen chloride
Tetrachloroethylene
indicating value 〔ppm〕
500 1000
Coexistence
Interference
0
200
100
1,2-Dichloroethylene〔ppm〕�
FIG.2 Influence of 1,2-Dichloroethylene
Tetrachloroethylene
indicating value 〔ppm〕
50 100
Coexistence
Interference
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
150 - 172 150 138 134
100 144 116 100 92 88
50 70 56 50 46 44
30 40 36 30 28 26
20 22 21 20 19 18
10 10 10 10 10 10