产品列表PRODUCTS LIST
180S DICHLOROMETHANE 二氯甲烷CH2Cl2气体检测管 Dichloromethane 10-1000ppm, 总代理销售日本光明理化学工业株式会社北川式气体检测管,提供日本光明理化学工业株式会社北川式气体检测管介绍,日本光明理化学工业株式会社北川式气体检测管应用,日本光明理化学工业株式会社北川式气体检测管价格,日本光明理化学工业株式会社北川式气体检测管说明书,日本光明理化学工业株式会社北川式气体检测管售后服务,日本光明理化学工业株式会社北川式气体检测管中国总代理
更新时间:2024-10-30
在线留言二氯甲烷CH2Cl2气体检测管180S Dichloromethane 10-1000ppm
Tube No.
180S
1. PERFORMANCE
1) Measuring range :30-1,000 ppm 10-200 ppm
Number of pump strokes 2(200ml) 4(400ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :5 ppm(400ml)
4) Shelf life :2 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :5~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 2 pump strokes
8) Colour change :White→Reddish orange
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high:15%
3. CHEMICAL REACTION
Chlorine is produced an Oxidizer. By reacting between this Chlorine and o-Toluidine, Orthoquinone is produced.
CH3OCI3+CrO3+H2SO4→CI2
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
DETECTOR TUBE
(Olive)
PRETREAT TUBE
1000
800
500
400
200
100
30
ppm
200ml
180S DICHLOROMETHANE
Substance Interference Coexistence
Halogens or Halogenated hydrocarbons Similar stain is produced. Higher readings are given.
FIG.1
(NOTE)
When the concentration is below 200 ppm, 4 pump strokes can be used to determine the lower concentration
Following formula is available for the actual concentration.
Actual concentration=1/3×Temperature corrected value.
0
200
400
600
800
1000
Chlorine〔ppm〕�
FIG.1 Influence of Chlorine