产品列表PRODUCTS LIST

首页>产品中心>日本光明北川式气体检测管>气体快速检测管>111U乙酸乙酯气体检测管111U ETHYL ACETATE 20-300ppm

乙酸乙酯气体检测管111U ETHYL ACETATE 20-300ppm

简要描述:

乙酸乙酯C6H10 111U ETHYL ACETATE (+) Cyclohexene 环己烯 20-300ppm,重要的是,不同的浓度意味着被抽取的空气中的氧气的量不同,当有大量氧气时(浓度较高)氧气分子会前行的很远(变白的区域很长)。然而当只有很少的氧气时(浓度较低),就会很快穿过试剂与之反应而耗尽,这样变白的区域就很短。

更新时间:2024-10-06

发邮件给我们:554328895@qq.com

分享到: 1
在线留言
乙酸乙酯气体检测管111U ETHYL ACETATE 20-300ppm

乙酸乙酯气体检测管111U ETHYL ACETATE 20-300ppm

Tube No.
111U
1. PERFORMANCE
1) Measuring range :20-1,000 ppm
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :10 ppm
4) Shelf life : :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH3CO2C2H5+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
1000
800
600
400
200
100
50
20
􀀀􀀀􀀀􀀀􀀀

􀀀􀀀
􀀀 􀀀
111U ETHYL ACETATE
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 10℃ 15℃ 20℃ 25℃ 30℃ 35℃ 40℃
(ppm) (50°F) (59°F) (68°F) (77°F) (86°F) (95°F) (104°F)
1,000 1,330 1,200 1,000 830 700 600 500
800 1,080 950 800 670 550 470 400
600 830 720 600 500 420 380 300
400 570 480 400 330 280 240 200
200 280 240 200 170 140 120 100
100 150 120 100 80 70 60 50
50 70 60 50 40 30 20 10
20 30 25 20 15 10 7 4
Substance Interference Coexistence
Alcohols Similar stain is produced. Higher reading are given.
Esters 〃〃
Ketones 〃〃
Aromatic hydrocarbons 〃〃
Whole reagent is discoloured to It the top of brown stain is clear,
Paraffin hydrocarbons Pale brown. athffee catcecdu.racy of readings is not
Halogenated hydrocarbons FIG.1 〃〃
0
200
400
600
800
1000
Trichloroethylene〔ppm〕�
FIG.1 Influence of Trichloroethylene
Ethyl acetate indicating value 〔ppm〕
100 200 300
112
Tube No.
1. PERFORMANCE
1) Measuring range :5-60 ppm
Number of pump strokes 2(200ml)
2) Sampling time :3 minutes/2 pump strokes
3) Detectable limit :0.5 ppm
4) Shelf life : :2 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Graduations printed on the tube are calibrated by Methyl acrylate at 2 pump strokes
and Ethyl acrylate concentration is determined by using a conversion chart.
8) Colour change :Yellow→Pale blue
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH2=CHCO2CH2CH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
60
40
20
10
5
ppm
200ml
DEHUMIDIFIER
(White)
􀀀􀀀􀀀􀀀􀀀

􀀀􀀀
211U C ETHYL ACR